1. General Introduction

YTA70P temperature transmitter owns almost complete same specifiation as YTA70, unique compact rail mount details which enable to be mounted in a cubic panels, saving mounting spaces.

2. Sensors inputs, measure range and accuracy

Sensor Type

Standard

Measure range

Minimum Span

Accuracy

 

Temperature effects/10

T/C

B

IEC60584

-400 to 1820

752 to 3308

200

360

±0.1% or ±1℃

±0.05% or ±1℃

E

-100 to 1000

-148 to 1832

50

90

±0.1% or ±0.5℃

±0.05% or ±0.25℃

J

-100 to 1200

-148 to 2192

50

90

K

-180 to 1372

-292 to 2502

50

90

N

-180 to 1300

-292 to 2372

100

180

R

-50 to 1760

-58 to 3200

200

360

±0.1% or ±1℃

±0.05% or ±1℃

S

-50 to 1760

-58 to 3200

200

360

T

-200 to 400

-328 to 752

50

90

±0.1% or ±0.5℃

±0.05% or ±0.25℃

L

DIN43710

-100 to 900

-148 to 1652

50

90

U

-200 to 600

-328 to 1112

75

135

Lr

Gost 3044-84

-200 to 800

-328 to 1472

50

90

±0.1% or ±1℃

±0.05% or ±1℃

W3

ASTM E988-90

0 to 2300

32 to 4172

200

360

W5

0 to 2300

32 to 4172

200

360

RTD

PT100

IEC60751

-200 to 850

-328 to 1562

10

18

±0.1% or ±0.1℃

±0.05% or ±0.05℃

Ni100

DIN43760

-60 to 250

-76 to 482

10

18

±0.1% or ±0.2℃

±0.05% or ±0.05℃

mV

-800 to 800 mV

±2.5 mV

±0.1% or ±0.01mV

±0.05% or ±0.05mV

ohm

0 to 7000 Ω

±25 Ω

±0.1% or ±0.1Ω

±0.05% or ±0.05Ω

3. Model and suffix codes

Main Code

Suffix Codes

Description

YTA70P




Temperature Transmitter   (Panel Mount Type)

Output Signal

-J



4 to 20mA DC with   digital communication (HART 5/HART 7 protocol)


A


Always A

Optional codes



/V2S

ATEX, FM, IECEx, and   CSA Intrinsically safe Approval